Welcome to the Trackit SR Optical Tracking 3D Scanner User Manual.
This manual provides comprehensive guidance for the installation, connection, calibration, maintenance, and operation of the Trackit SR 3D scanning system. It is designed to help you quickly get started and effectively use your scanner.





Trackit SR includes Revo Track, which integrates scanning, post-processing, measurement, and model alignment into a single workflow for efficient 3D data acquisition and processing.
This section provides an overview of the complete Trackit SR scanning workflow.
Pull the tripod legs outward until they lock into place. (A click sound indicates it has locked into one of the three preset positions shown in descending order.) Adjust the center column and the legs to your required height.
Note:Ensure that the legs stay at the same preset position.
Step 1: Align the central screw hole of the Quick Release Clamp with the screw on the tripod, and rotate the clamp until it is tightened.

Step 2: Slide the Quick Release Plate on the bottom of the Tracker into the groove of the Quick Release Clamp, and ensure the white lines on the front are aligned. Then, tighten the locking knob clockwise.
-install the tracker.png)
Step 1: Connect the Power Adapter’s DC end to Tracker’s “DC” port, and the AC end to a power supply.
-connect_the_power_source.png)
Step 2: Connect the PC Cable’s L-shaped end to the Tracker’s “PC” port and tighten the screws. Insert the straight end into a USB-C port on your PC.
-connect_the_pc.png)
Step 3: Connect the Scanner Cable’s L-shaped end to the Tracker’s “12V SCAN” port and tighten the screws. Connect the straight end to the USB-C port on the Scanner and tighten the screws.
-connect_the_scanner(1).png)
Step 4: Turn on the power switch on the Scanner. Wait a few seconds until the Scanner’s indicator light turns solid green.

Step 5: Open Revo Track. When both the Scanner and Tracker appear as connected in the software, the system is ready.

Before using Wi-Fi connection for the first time, connect the Trackit SR via a wired connection to configure the Wi-Fi settings in Revo Track.
Step 1: Follow the steps in "Wired Connection" to connect the device. Then open the software and go to Scanner Settings > Wi-Fi Settings.

Step 2: In Wi-Fi settings, ensure "Host" is selected, then click "Confirm". The software will send the Wi-Fi configuration to the Tracker and Scanner.

Step 3: Open the Scanner’s battery holder cover, insert the battery according to the indicated polarity, and close the cover.

Step 4: Turn on the power switch on the Scanner. Wait a few seconds until the Scanner’s indicator light turns solid green.

Step 5: Align the screw on top of the Battery Grip with the threaded hole on the side of the tripod, then turn the Battery Grip clockwise until securely tightened.
Note: The Battery Grip is not included in the standard edition. You can purchase it from Revopoint official online store.

Step 6: Connect the Battery Grip to the “DC” port on the back of the Tracker using the Battery Grip Cable.
-connect_the_battery_grip.png)
Step 7:Open your computer's Wi-Fi list, search for the Wi-Fi network named TrackitSR-Scanner-REVO-XXXXXXXX, enter the default password “Revopoint3d”, and connect.

Step 8: Open Revo Track. When both the Scanner and Tracker appear as connected in the software, the system is ready.

Before starting calibration, Revo Track allows you to check the device accuracy to determine whether calibration is necessary.
① On the lower-left corner of the software home page, click Device Accuracy Check to enter the accuracy check interface.

② Scan the QR code on the back of the Calibration Board with the Scanner, then click "Download and Import Online" to import the calibration information.

③ Place the Calibration Board flat on the table with the side markers facing the Tracker. Then move the Tracker as instructed until it reaches the on-screen guide position. The software will automatically enter the calibration page.
.png)
④ Hold the Scanner facing the Calibration Board, and rotate it as instructed on the screen to complete the accuracy check.
Note:When holding the Scanner, avoid obstructing the Tracker's FOV. Ensure that the Tracker maintains a clear view of both the Scanner and the Calibration Board.

1) Enter the Calibration Process
If the accuracy check fails, click Device Calibration to proceed to calibration. For the initial calibration, it is recommended to perform full calibration.
Note: After entering the calibration process, the software will automatically warm up the device.
-full calibration-a.png)
2) Import SN Info
Follow the on-screen prompts to scan the QR codes on the Calibration Pole and the Calibration Board, then click “Download and Import Online” to import their SN info.

3) Start Scanner Calibration
Place the Calibration Board flat on the table with the white side facing up. Hold the Scanner with its cameras facing the board. Follow the on-screen prompts to adjust the Scanner’s tilt, pitch, and height until calibration in all directions is completed.

After Scanner calibration, the software calculates the results. If qualified, click "Continue" to proceed to the Tracker accuracy check and calibration.
1) Install the Quick Release Clamp
Extend the tripod’s center column to its maximum height. Align the central screw hole of the Quick Release Clamp with the screw on the tripod, and rotate the clamp until it’s tightened.
-install the quick release clamp.jpg)
2) Install the Gimbal Adapter Plate
Slide the Gimbal Adapter Plate into the groove of the Quick Release Clamp, and ensure the white lines on the front are aligned. Then, tighten the locking knob.
-install the adapter plate.jpg)
3) Install the Gimbal
Keep the gimbal’s USB-C port facing upward, and secure the gimbal to the adapter plate using the supplied screw.
-install the gimbal-a.jpg)
4) Install the Calibration Pole
Keep the Calibration Pole’s Up arrow facing upward. Align the pole’s screw hole with the screw on the gimbal, then tighten the thumbwheel until the pole is securely fastened.
-install the calibration pole-a.jpg)
5) Connect the Gimbal to the Tracker
Connect the gimbal to the gimbal port on the back of the Tracker using the gimbal cable (yellow cable) and tighten the screws.
-connect the gimbal-a.jpg)
6) Tracker Calibration
Hold the Tracker with its cameras facing the Calibration Pole. Then move the Tracker according to the software prompts until it is aligned with the indicator frame on the interface. Perform the Tracker calibration at different distances.
--trcaker calibration.jpg)
1) Install the Tracker
Remove the Calibration Pole, dual-axis gimbal, and Gimbal Adapter Plate from the tripod, and mount the Tracker onto the Quick Release Clamp.
-install the tracker.png)
2) Place the Calibration Board
Place the Calibration Board flat on the table with the black side facing up and the marker side facing the Tracker. Move the Tracker according to the on-screen prompts until the board is aligned with the indicator frame on the interface.
-place the calibration board.png)
3) Hand-eye Calibration
Hold the Scanner and rotate it to the designated angles as instructed on the screen. Perform calibration for each pose in the left-right and front-back directions. Repeat the above steps until the hand-eye calibration is completed for all required angles.
Note:When holding the Scanner, avoid obstructing the Tracker's FOV. Ensure that the Tracker maintains a clear view of both the Scanner and the Calibration Board.
.png)
Trackit SR supports three scanning modes: Track Scan, Multi-position Scan, and Marker Scan. This section compares the three scanning modes and helps you choose the appropriate mode for your scanning task. Detailed software operations are available in the Revo Track User Manual.